Your purchase will NOT include: This product is pre-owned and may not include the same accessories as a new, unused unit. - other accessories not shown or listed here.Prior to testing this autoclave, the Reservoir Tank was cleaned and flushed, and a new Condenser Coil, Air Jet Valve and Safety Valve were installed. Installed a new Power Switch, Control Thermostat, Multi-Purpose Valve and Door Cover. Cleaned the sterilizing chamber and installed a new Door Seal and Door Bellows Assembly. Also cleaned the sterilizing trays and rack. Then proceeded to test this autoclave by levelling it for proper chamber filling, connecting it to a 120 volt power source and adding distilled water. Than ran sterilizing cycle of 250 F for 30 minutes and 270 F for 15 minutes several times. To verify that this sterilizer is achieving it's maximum required sterilization temperature during a cycle, it was tested using a Max Register Thermometer, and found to exceed 270 F for an entire cycle.
Technician Testing Notes
Prior to testing this autoclave, the Reservoir Tank was cleaned and flushed, and a new Condenser Coil, Air Jet Valve and Safety Valve were installed. Installed a new Power Switch, Control Thermostat, Multi-Purpose Valve and Door Cover. Cleaned the sterilizing chamber and installed a new Door Seal and Door Bellows Assembly. Also cleaned the sterilizing trays and rack. Then proceeded to test this autoclave by levelling it for proper chamber filling, connecting it to a 120 volt power source and adding distilled water. Than ran sterilizing cycle of 250 F for 30 minutes and 270 F for 15 minutes several times. To verify that this sterilizer is achieving it's maximum required sterilization temperature during a cycle, it was tested using a Max Register Thermometer, and found to exceed 270 F for an entire cycle.
Prior to testing this autoclave, the Reservoir Tank was cleaned and flushed, and a new Condenser Coil, Air Jet Valve and Safety Valve were installed. Installed a new Power Switch, Control Thermostat, Multi-Purpose Valve and Door Cover. Cleaned the sterilizing chamber and installed a new Door Seal and Door Bellows Assembly. Also cleaned the sterilizing trays and rack. Then proceeded to test this autoclave by levelling it for proper chamber filling, connecting it to a 120 volt power source and adding distilled water. Than ran sterilizing cycle of 250 F for 30 minutes and 270 F for 15 minutes several times. To verify that this sterilizer is achieving it's maximum required sterilization temperature during a cycle, it was tested using a Max Register Thermometer, and found to exceed 270 F for an entire cycle.
Product Condition
- Manufacturer: Tuttnauer
- Model: 2340M
- Serial Number: 2609844
- Power Requirements: 120V
- Year: 2007
- Sterilizer Type: steam chamber
- Timer Function: features a basic timer with one timing option
- Controls: analog control knobs
- Chamber Depth: 40 cm
- Chamber Diameter: 23 cm
- Number of Trays: 3 trays
- Cycle Count: Cycle count is not available on manual autoclaves
Item Includes:
1 - Tuttnauer 2340M Manual Autoclave - Steam Sterilizer 3 - Sterilizing Trays w/ Rack 1 - Drain Hose 1 - Power Cord
Item Excludes:
This product is pre-owned and may not include the same accessories as a new, unused unit.
1 - Tuttnauer 2340M Manual Autoclave - Steam Sterilizer 3 - Sterilizing Trays w/ Rack 1 - Drain Hose 1 - Power Cord
Item Excludes:
This product is pre-owned and may not include the same accessories as a new, unused unit.
Shipping calculated at checkout.
30 Day Return Guarantee
Your satisfaction is important to us: Have your equipment installed by a factory-trained technician or Atlas-approved vendor to qualify for our return policy.See our Terms and Conditions of Sale for details.