Your purchase will NOT include: This product is pre-owned and may not include the same accessories as a new, unused unit. - other accessories not shown or listed here.Before this autoclave was tested, the Reservoir Tank was cleaned and flushed and a new Air Jet Valve, Safety Valve and Reservoir Cap were installed. The Fill Valve Solenoid, Air Valve Solenoid, Exhaust Valve Solenoid and Dump Valve Solenoid were Replaced. Installed a new Door Gasket and Door Bellows Assembly. Installed a new Air Drying Pump, HEPA Air Drying Filter and Cap. Cleaned the sterilizing trays and rack. To test this sterilizer, it was connected to a 120 volt power source, distilled water was added to the reservoir and the Pressure and Temperature Sensors were calibrated. Then it was run through all it's preset sterilizing and drying cycles. To verify that it is achieving it's maximum required sterilizing cycle during a cycle, it was tested using a Max Register Thermometer, and found to exceed 273 F during a cycle.
Technician Testing Notes
Before this autoclave was tested, the Reservoir Tank was cleaned and flushed and a new Air Jet Valve, Safety Valve and Reservoir Cap were installed. The Fill Valve Solenoid, Air Valve Solenoid, Exhaust Valve Solenoid and Dump Valve Solenoid were Replaced. Installed a new Door Gasket and Door Bellows Assembly. Installed a new Air Drying Pump, HEPA Air Drying Filter and Cap. Cleaned the sterilizing trays and rack. To test this sterilizer, it was connected to a 120 volt power source, distilled water was added to the reservoir and the Pressure and Temperature Sensors were calibrated. Then it was run through all it's preset sterilizing and drying cycles. To verify that it is achieving it's maximum required sterilizing cycle during a cycle, it was tested using a Max Register Thermometer, and found to exceed 273 F during a cycle.
Before this autoclave was tested, the Reservoir Tank was cleaned and flushed and a new Air Jet Valve, Safety Valve and Reservoir Cap were installed. The Fill Valve Solenoid, Air Valve Solenoid, Exhaust Valve Solenoid and Dump Valve Solenoid were Replaced. Installed a new Door Gasket and Door Bellows Assembly. Installed a new Air Drying Pump, HEPA Air Drying Filter and Cap. Cleaned the sterilizing trays and rack. To test this sterilizer, it was connected to a 120 volt power source, distilled water was added to the reservoir and the Pressure and Temperature Sensors were calibrated. Then it was run through all it's preset sterilizing and drying cycles. To verify that it is achieving it's maximum required sterilizing cycle during a cycle, it was tested using a Max Register Thermometer, and found to exceed 273 F during a cycle.
Product Condition
- Manufacturer: Tuttnauer
- Model: EZ9
- Serial Number: 2408838
- Power Requirements: 120V
- Year: 2005
- Sterilizer Type: steam chamber
- Timer Function: features a timer with a variety of cycle options
- Controls: digital control display
- Chamber Depth: 40 cm
- Chamber Diameter: 23 cm
- Number of Trays: 3 trays
- Cycle Count: 4,200
Item Includes:
1 - Tuttnauer EZ9 Autoclave Steam Sterilizer 3 - Sterilizing Trays w/Rack 1 - Tray handle 1 - Drain Hose 1 - Power Cord
Item Excludes:
This product is pre-owned and may not include the same accessories as a new, unused unit.
1 - Tuttnauer EZ9 Autoclave Steam Sterilizer 3 - Sterilizing Trays w/Rack 1 - Tray handle 1 - Drain Hose 1 - Power Cord
Item Excludes:
This product is pre-owned and may not include the same accessories as a new, unused unit.
Shipping calculated at checkout.
30 Day Return Guarantee
Your satisfaction is important to us: Have your equipment installed by a factory-trained technician or Atlas-approved vendor to qualify for our return policy.See our Terms and Conditions of Sale for details.