SKU: 113247
Condition report
Condition
MARKS & STAINS
Known issues: Damaged muffle, cracked screen, damaged firing plateYellow staining
Test notes
Unit was connected to power and a tester vacuum pump. The unit passed its self‑checks and diagnostic testing, including keypad, vacuum, and heat tests, but did not pass the muffle test. During the display test a crack was observed on the screen, and the firing plate was found to be cracked in several places. This unit is to be listed as a parts unit.
Specifications
ManufacturerIvoclar Vivadent
ModelEP 600
Serial Number80316
Power Requirements120V
Maximum Temperature1075C
Vacuum Pump StatusNot included
Hours of Use5976
Furnace TypePressing
Chamber ShapeRound
Chamber Dimensions100mm x 90mm
Software Version4.0.0
Warranty
No returns accepted
Shipping & returns
Shipping (US)
| Ship weight | Rate |
|---|---|
| Under 50 lb | Free ground parcel |
| 50–100 lb | $29 ground parcel |
| Freight items | $299 per freight item · ground parcel free with freight |
Ships from Boise, ID.
Shipping calculated at checkout.
Need your order to deliver by a specific date? Contact our team for a guaranteed delivery quote.
Returns
30 Day Return Guarantee
$750.00
Processing fee (3%) at checkout · Waived for ACH