Condition report
Condition
Scratches
Test notes
Performed a visual inspection of the unit to check for physical damage and missing parts.
Connected the unit to the N₂O and O₂ gas supplies and checked all connections to ensure there were no gas leaks.
Checked the regulators to verify they held the correct operating pressures (PSI).
Tested the balance between the nitrous oxide and oxygen supplies to verify proper unit operation.
Tested the oxygen failure safety switch by cutting off the oxygen supply and verifying that the safety switch activated as intended.
Observed the unit during operation and noted chattering from one regulator.
Confirmed that the unit functioned during the tests performed, with regulator chattering documented as an observed issue.
Specifications
Warranty
30 Day Return Guarantee
Shipping & returns
Shipping (US)
| Ship weight | Rate |
|---|---|
| Under 50 lb | Free ground parcel |
| 50–100 lb | $29 ground parcel |
| Freight items | $299 per freight item · ground parcel free with freight |
Ships from Boise, ID.
Need your order to deliver by a specific date? Contact our team for a guaranteed delivery quote.
Returns
30 Day Return Guarantee
Your satisfaction is important to us: Have your equipment installed by a factory-trained technician or Atlas-approved vendor to qualify for our return policy.See our Terms and Conditions of Sale for details.
Processing fee (3%) at checkout · Waived for ACH