Condition report
Condition
MARKS & SCRATCHES
Test notes
Unit was visually inspected for missing components and damages. Unit was connected to an acquisition unit and to 120V power. Unit ran its start up sequence with no issues or errors. The unit was put through a calibration cycle followed by a stepper motor, force sensor, and a continuous load cycle to verify units functionality. Unit passed all tests/cycles without any issues or errors. Unit is fully functional.