Condition report
Condition
Test notes
This autoclave is 2018 Model with 200 cycles run it. A diagnostic test was preformed and all the major components were found to be functioning. A a preventive measure, the Pressure Relief Valve was replaced. Also the front plastic fascia cover was replaced, and the sterilizing tray was cleaned.
THIS AUTOCLAVE REQUIRES A 230 VOLT A/C POWER OUTLET TO FUNCTION. PLEASE NOTE THIS SPECIFICATION BEFORE PURCHASING.
To test this autoclave, it was first connected to a 230 volt power source, the included condenser waste bottle was attached, distilled water was added to the sterilizer's reservoir tank, and the tray was inserted. Then the autoclave was run through all it's preset sterilizing and drying cycles several times. The sterilizer was found to be leak free and functioning to spec during it's testing. To verify that this autoclave is making it's maximum required sterilizing temperature, it was tested using a Max Register Thermometer, and found to exceed 273 F during a sterilizing cycle.