Your purchase will NOT include: This product is pre-owned and may not include the same accessories as a new, unused unit. - other accessories not shown or listed here.This autoclave is a 2018 model with an unknown number of cycles run on it, due to it being a manual model autoclave. Initial testing of this autoclave reveled that it had numerus issues that needed attention before the sterilizer could be tested. All the known issues were taken care of, which included replacing the Door Switch, Power Switch and Multi-Purpose Valve. Then the Reservoir Tank was cleaned and flushed and a new Safety Valve, Condenser Coil, Air Jet Valve and Reservoir Cap were installed. Cleaned the sterilizing chamber and installed a new Door Gasket and Door Bellows Assembly. Cleaned the sterilizing trays and rack. To test this autoclave, it was first connected to a 120 volt power source, distilled water was added, it was then properly levelled as per it's chamber size requirement and then testing began. This sterilizer was run through cycles of 270 F for 10 minutes and 250 for 30 minutes several times. To verify that this autoclave is achieving it's maximum required sterilizing temperature during a cycle, it was temperature tested with the use of a Max Register Thermometer, and found to exceed 270 F during a sterilization cycle.
Technician Testing Notes
This autoclave is a 2018 model with an unknown number of cycles run on it, due to it being a manual model autoclave. Initial testing of this autoclave reveled that it had numerus issues that needed attention before the sterilizer could be tested. All the known issues were taken care of, which included replacing the Door Switch, Power Switch and Multi-Purpose Valve. Then the Reservoir Tank was cleaned and flushed and a new Safety Valve, Condenser Coil, Air Jet Valve and Reservoir Cap were installed. Cleaned the sterilizing chamber and installed a new Door Gasket and Door Bellows Assembly. Cleaned the sterilizing trays and rack. To test this autoclave, it was first connected to a 120 volt power source, distilled water was added, it was then properly levelled as per it's chamber size requirement and then testing began. This sterilizer was run through cycles of 270 F for 10 minutes and 250 for 30 minutes several times. To verify that this autoclave is achieving it's maximum required sterilizing temperature during a cycle, it was temperature tested with the use of a Max Register Thermometer, and found to exceed 270 F during a sterilization cycle.
This autoclave is a 2018 model with an unknown number of cycles run on it, due to it being a manual model autoclave. Initial testing of this autoclave reveled that it had numerus issues that needed attention before the sterilizer could be tested. All the known issues were taken care of, which included replacing the Door Switch, Power Switch and Multi-Purpose Valve. Then the Reservoir Tank was cleaned and flushed and a new Safety Valve, Condenser Coil, Air Jet Valve and Reservoir Cap were installed. Cleaned the sterilizing chamber and installed a new Door Gasket and Door Bellows Assembly. Cleaned the sterilizing trays and rack. To test this autoclave, it was first connected to a 120 volt power source, distilled water was added, it was then properly levelled as per it's chamber size requirement and then testing began. This sterilizer was run through cycles of 270 F for 10 minutes and 250 for 30 minutes several times. To verify that this autoclave is achieving it's maximum required sterilizing temperature during a cycle, it was temperature tested with the use of a Max Register Thermometer, and found to exceed 270 F during a sterilization cycle.
Product Condition
- Manufacturer: Tuttnauer
- Model: 2540M
- Serial Number: 18030985
- Power Requirements: 120V
- Year: 2018
- Sterilizer Type: steam chamber
- Timer Function: features a timer with a variety of cycle options
- Controls: analog control knobs
- Chamber Depth: 40 cm
- Chamber Diameter: 25 cm
- Number of Trays: 4 trays
- Cycle Count: Cycle counts are not available on manual autoclave models.
Item Includes:
1 - Tuttnauer 2540M Manual Autoclave Steam Sterilizer 4 - Sterilizing Trays w/ Rack 1 - Drain Hose 1 - Power Cord
Item Excludes:
This product is pre-owned and may not include the same accessories as a new, unused unit.
1 - Tuttnauer 2540M Manual Autoclave Steam Sterilizer 4 - Sterilizing Trays w/ Rack 1 - Drain Hose 1 - Power Cord
Item Excludes:
This product is pre-owned and may not include the same accessories as a new, unused unit.
Shipping calculated at checkout.
Need your order to deliver by a specific date? Contact our team for a guaranteed delivery quote.
30 Day Return Guarantee
Your satisfaction is important to us: Have your equipment installed by a factory-trained technician or Atlas-approved vendor to qualify for our return policy.See our Terms and Conditions of Sale for details.