Condition report
Condition
MARKS & STAINS
Test notes
This autoclave is a 2006 model with 3,300 run in it. It was inspected inside and out for obvious defects, then it was put through an initial functionality test to verify it is in working order. Then it semi-disassembled to clean. To insure this sterilizers continued operation, preventative maintenance parts were installed, which include, a new Door Switch, Power Switch, Door Seal, Fill Valve, Dump Valve, Air Valve and Exhaust Valve. The Reservoir Tank was cleaned and flushed, and a new Reservoir Cap, Condenser Coil, Safety Valve, and Air Jet Valve were installed. The Chamber and Water Fill Filter were cleaned and a new HEPA Air Drying Filter and Cap were installed. Cleaned the Sterilizing Tray and Rack.
Then testing proceeded on this sterilizer. Connected it to a 120 volt power source, added distilled water and ran the autoclave through all it's preset sterilizing and drying cycles. To verify that this autoclave is achieving it's required maximum sterilizing temperature, it was temperature tested using a Max Register Thermometer, and found to reach 273 F for an entire cycle.
Specifications
Warranty
30 Day Return Guarantee
Shipping & returns
Shipping (US)
| Ship weight | Rate |
|---|---|
| Under 50 lb | Free ground parcel |
| 50–100 lb | $29 ground parcel |
| Freight items | $299 per freight item · ground parcel free with freight |
Ships from Boise, ID.
Need your order to deliver by a specific date? Contact our team for a guaranteed delivery quote.
Returns
30 Day Return Guarantee
Your satisfaction is important to us: Have your equipment installed by a factory-trained technician or Atlas-approved vendor to qualify for our return policy.See our Terms and Conditions of Sale for details.
Processing fee (3%) at checkout · Waived for ACH